Installation

ICP-MS Replacement and Water Chiller Manifold Rebuild for Restored Tuning Performance

A laboratory contacted us for ICP-MS autotune failure and plasma ignition issues after maintenance was performed on an older system. During diagnostics, the RF work coil was found contacting the bonnet, which had caused misalignment in the torch assembly. The torch components, gas settings, RF supply output, RF coil, and related connections were inspected and tested, but the instrument continued to have ignition problems when correctly aligned. Because the next likely repair path involved RF gen...
Read More

Caron Environmental Chamber Installation and Drainage System Repair

When a customer needs Caron environmental chamber installation and drainage setup, especially for pre-owned units, hidden infrastructure issues can prevent proper operation from the start. In this case, multiple chambers were delivered in unknown condition and required full installation, testing, and system integration. During initial inspection, the chambers were found to be missing proper provisions for external drain pumps, and the existing water supply and drainage configuration was not comp...
Read More

Agilent 1100 Pressure Problem Repair

Agilent 1100 pressure problem repair was needed for a system that would not build pressure during startup after installation and connection. The HPLC repeatedly failed the pressure test even though the leak test passed, which prevented the system from moving into normal operation. Our diagnostics included connection review, firmware updates, pump service, and testing of the related support equipment. The pump seals and pump head components were serviced first, but the pressure problem remained....
Read More

Resolving a PerkinElmer EnVision Software Issue

This PerkinElmer EnVision software issue appeared after the system workstation was removed from a previous domain environment. The customer needed the instrument restored for standalone operation, but the existing computer setup was preventing the EnVision software from functioning correctly and blocking normal communication with the instrument. Our diagnostics showed that the problem was caused by leftover domain-related and Windows configuration settings from the previous installation. A softw...
Read More